Intermittent output failure of a VVI device due to the disintegration of the generator

Citation
A. Bohm et al., Intermittent output failure of a VVI device due to the disintegration of the generator, PACE, 24(1), 2001, pp. 127-128
Citations number
5
Language
INGLESE
art.tipo
Article
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
0147-8389 → ACNP
Volume
24
Issue
1
Year of publication
2001
Pages
127 - 128
Database
ISI
SICI code
0147-8389(200101)24:1<127:IOFOAV>2.0.ZU;2-Y
Abstract
The metal case and the header of a Biovallees Coralite 247 pacemaker fell a part in a patient resulting in syncope due to myopotential inhibition. Upon recalling our patients with that type of device, we found a similar techno logical failure in 12 of 16 of them. This finding warrants a consideration for recalling that Biovallees device.