Capillary zone electrophoresis was employed for the determination of reserp in using an end-column amperometric detection with a carbon fiber micro-dis k array electrode at a constant potential of +1.30 V vs, saturated calomel electrode. The optimum conditions of separation and detection are 1.0 x 10( -4) mol/l Tris-8.4 x 10(-5) mol/l HCl for the buffer solution, 20 kV for th e separation voltage, 5 kV and 10s for the injection voltage and the inject ion time, respectively. The limit of detection is 8.0 x 10(-7) mol/l or 1.4 fmol (S/N=2). The relative standard deviation is 2.5% for the migration ti me and 1.5% for the electrophoretic peak current. The method was applied to the determination of reserpin in human serum. (C) 2000 Elsevier Science B. V. All rights reserved.