The mechanical design of the EUROBALL neutron wall and neutron detectors, a
nd their performance measured with a Cm-246,Cm-248 fission source are descr
ibed. The array consists of 15 pseudohexaconical detector units subdivided
into three, 149 mm high, hermetically separated segments and a smaller cent
ral pentagonal unit subdivided into five segments. The detectors are filled
with Bicron BC501A liquid scintillator. Each section of the hexaconical de
tectors is viewed by a 130 mm diameter Philips XP4512PA photomultiplier whi
le the sections of pentagonal detectors are viewed by Philips XP4312B PMTs.
The tests of n-gamma discrimination performed by zero-crossing and time-of
-flight methods show a full separation of gamma- and neutron events down to
50 keV recoil electron energy. These tests demonstrate the excellent timin
g properties of the detectors and an average time resolution of 1.56 ns. Th
e factors determining the efficiency of neutron detectors are discussed. Th
e total efficiency for the full array for a symmetric fusion-evaporation re
action is predicted to be 0.30. (C) 1999 Elsevier Science B.V. All rights r
eserved.