T. Palenciadominguez et al., EVALUATION OF A NEW MEASUREMENT PROCEDURE FOR THE CONCENTRATION OF FERRITIN IN SERUM, European journal of clinical chemistry and clinical biochemistry, 35(2), 1997, pp. 117-120
The aim of this study was the evaluation of an immunoturbidimetric mea
surement procedure for the concentration of ferritin in serum, based o
n the new Tina-quant Ferritin reagents kit (Boehringer Mannheim) in wh
ich the antibody-coated latex particles have been modified with regard
to the previous Tina-quant Ferritin reagents kit. The evaluation was
carried out using the BM/Hitachi 917 Analyzer. The evaluation included
the within-run and between-day imprecisions estimation, the compariso
n of measurement procedures and the assessment of the measuring range.
An additional study of between-day imprecision was done using 4 mu l
and 20 mu l of sample. Detection limit was studied using BM/Hitachi 91
7 (4 mu l and 20 mu l of sample), ES 300 and Cobas-Core analyzers. The
coefficients of variation obtained in the between-day imprecision stu
dy are lower when 20 mu l of sample is used instead of 4 mu l. In the
measurement procedure using 20 mu l, the coefficient of variation obse
rved at physiological concentrations is 2.4%. The detection limit with
20 mu l of sample is lower than with 4 mu l. Therefore, in ferropenic
anaemia studies it is convenient to use the measurement procedure wit
h 20 mu l sample volumes, since it possesses the best metrological cha
racteristics for this purpose.